Difference between revisions of "STEMI-2017/C2/Introduction-to-Kallows-Device/English"
From Script | Spoken-Tutorial
PoojaMoolya (Talk | contribs) (Created page with "<div style="margin-left:-0.035cm;margin-right:0cm;">Topic: KALLOWS - Positioning ECG leads, BP cuff and SpO2 probe</div> <div style="margin-left:-0.035cm;margin-right:0cm;">C...") |
PoojaMoolya (Talk | contribs) |
||
Line 99: | Line 99: | ||
|| | || | ||
|| '''Pls note: ''' | || '''Pls note: ''' | ||
− | * <div style="margin-left:1.27cm;margin-right:0cm;">Reusable sensors may be used on the same site for a maximum period of 4 hrs, | + | * <div style="margin-left:1.27cm;margin-right:0cm;">Reusable sensors may be used on the same site for a maximum period of 4 hrs, Make sure that the site is inspected routinely to ensure skin integrity.</div> |
* <div style="margin-left:1.27cm;margin-right:0cm;">Do not use wet or damaged sensors. They may cause burns during the electro surgery or when applying other electrical device.</div> | * <div style="margin-left:1.27cm;margin-right:0cm;">Do not use wet or damaged sensors. They may cause burns during the electro surgery or when applying other electrical device.</div> | ||
* <div style="margin-left:1.27cm;margin-right:0cm;">Tissue damage/ burn can be caused by incorrect application or use of SpO2 sensor</div> | * <div style="margin-left:1.27cm;margin-right:0cm;">Tissue damage/ burn can be caused by incorrect application or use of SpO2 sensor</div> |
Revision as of 15:24, 29 June 2020
Topic: KALLOWS - Positioning ECG leads, BP cuff and SpO2 probe
Contributor Name: Jerry David, STEMI Team, Coimbatore
Reviewed by: Nancy Varkey, Spoken Tutorial Project, IIT Bombay
Keywords: STEMI Kit, NIBP, SpO2, ECG, Electrode, B.P cuff, Mobmon 12.0
VISUAL CUE | NARRATION | ||||||||||||||||||||||
Show Slide
Kallows STEMI Kit. |
Hello and welcome to this tutorial on the Kallows STEMI Kit. | ||||||||||||||||||||||
Learning Objectives | In this tutorial we will learn to -
| ||||||||||||||||||||||
Show Slide
Hardware Requirements |
To practise this tutorial, you will need the Kallows STEMI Kit. | ||||||||||||||||||||||
Show Image | The STEMI Kit consists of
| ||||||||||||||||||||||
Show image | This is the Mobmon device. | ||||||||||||||||||||||
Show image | It has a power button along with the charging port on the left side. | ||||||||||||||||||||||
Show image | And the SpO2 and ECG ports on the back side. | ||||||||||||||||||||||
Show image | This cable head has to be inserted into the ECG port on the Mobmon device. | ||||||||||||||||||||||
Show image | After connecting, secure the connection with the screws provided on either sides. | ||||||||||||||||||||||
Show image | Next, we will learn about the SpO2 probe. | ||||||||||||||||||||||
Show image | This probe is made up of the following parts-
| ||||||||||||||||||||||
Show image | Let’s now see how to position the SpO2 probe. | ||||||||||||||||||||||
Show image | Connect the oximetry probe/cable to the SpO2 connector on the Mobmon device. | ||||||||||||||||||||||
Show image | This is how it should look when connected. | ||||||||||||||||||||||
Show image | The patient's finger must be inserted right to the end of the sensor as shown in the figure. | ||||||||||||||||||||||
Show image | While selecting a sensor site, priority should be given to an extremity free of -
| ||||||||||||||||||||||
Show image | The typical sites for oximetry probe connections are -
| ||||||||||||||||||||||
Pls note:
| |||||||||||||||||||||||
NIBP CUFF PLACEMENT | Next, let’s learn how to position Blood Pressure cuff. | ||||||||||||||||||||||
Show image | Connect the cuff connector to NIBP Connector of the Bluetooth device. | ||||||||||||||||||||||
Show image |
| ||||||||||||||||||||||
Press the Start button to activate Bluetooth BP Monitor. | |||||||||||||||||||||||
Pls note:
Inaccurate measurements may be caused by incorrect application or use such as-
| |||||||||||||||||||||||
Do not connect the SpO2 sensor and the B.P cuff to the same limb of the patient.
| |||||||||||||||||||||||
Show image | Make sure that while measuring the NIBP, the tube of the cuff is not blocked or tangled.
| ||||||||||||||||||||||
Finally, let’s see how to position the ECG leads. | |||||||||||||||||||||||
PATIENT PREPARATION | Quality ECG rhythm can be obtained with good skin preparation and proper electrode placement. | ||||||||||||||||||||||
Follow the instructions given in the narration |
The following must be done for good patient preparation-
| ||||||||||||||||||||||
Show image | Note:
| ||||||||||||||||||||||
Show image | This is how the electrodes should be placed.
| ||||||||||||||||||||||
Make sure the power cord and patient cable do not cross each other. | |||||||||||||||||||||||
Show image | Switch on the Mobmon device by selecting power on/ off button on the side. | ||||||||||||||||||||||
STEMI App to be shown here | Select ECG tab on the STEMI device to view the ECG, in the ECG live steam page. | ||||||||||||||||||||||
Let us summarize. | |||||||||||||||||||||||
Let us summarize. | In this tutorial, we have learnt -
| ||||||||||||||||||||||
Show STEMI INDIA slide | STEMI INDIA
| ||||||||||||||||||||||
Show Spoken Tutorial Project Slide | Spoken Tutorial Project, IIT Bombay is funded by NMEICT, MHRD, Govt. of India.
| ||||||||||||||||||||||
Show THANK YOU slide | This tutorial has been contributed by
This is Nancy Varkey signing off. Thanks for joining. |